Good evening sir. I don't know if you'll be reading this. Today while doing ptca in an awmi, I ended up closing a flowing artery. Probably a dissection because of the workhorse wire . Was unable to cross with whisper wires of diffrent shapes. Switched to Gaia 2 that crossed and helped complete the case and restore flow. Watching your videos made it possible. Thanks 🙏🏼
@manosbrilakis
2 ай бұрын
Great to hear - glad things worked out well!
@farukakturk5388
3 ай бұрын
While pulling the entrapped microcath simultanously pushing it from the opposite site with a balloon or another Micro?
@manosbrilakis
2 ай бұрын
You do not want to push from the other side because the tips of microcatheters may get entangled and make things worse
@farukakturk5388
2 ай бұрын
@@manosbrilakis Can they? For retrograde cases? They would be on the same wire? But if you say so there is something that ı dd not consider.