#NOxBOX machine to give #nitric_oxide up to 80 ppm by #Bedfont Ltd..
Instagram Account : mohd_Saeed
SnapChat Account : Mohdalnadhri
Twitter Account : @R_Tmohammed
Zero Calibration
it should be done once we turn on the machine and it does not requires the circuit to be connected nor turning on of INO . it should be done Every 24 hrs while it is working on the patient , the INO Delivery will not be compromised .once you are in doubt about the proper function ability of the machine .
Calibration Circuit and System Test
the Connection and how it should be done you can see after 9 Minutes and 30 Seconds from starting the video . O2/INO Flow should be 10 LPM , and the INO Tanks should be opened . if it is passed , you Can set the Desired INO Dose to start the therapy and you Can see it after 13 minutes from starting the video .
Patient is Crashing and no time for Calibration
Turn on the Machine , let the Zero Calibration Successfully finished then Go Directly to SKIP TO DOSE option but make sure the Circuit is properly Connected and you Can see that after 14 minutes from starting the Video .NOXFLOW Sensor Connections can be seen after 4 minutes from starting the Video . after 14 Minutes you Can see patient Circuits with NoXFLOW Sensor and the Sampling Line .
STAND BY
Make the INO Dose Zero PPM . You Can See it after 22 minutes from starting the Video . it is Recommended to Shut-off the INO Cylinders and depressurized the INO Hoses.
Manual Bagging of the Patient .
We have to Know that we have Two Modes in this Machine Which are INTELLIGENT mode and MANUAL OVERRIDE MODE . You can See that after 25 Seconds from Starting the Video . to Do manual Bagging just go directly to 19 Minutes and 22 Seconds in the Video