Hello. 1. Do you know why I could not integrate all of the peaks? Can I integrate the peaks via manual integrator? (Though I could not find where this function) 2. Should I fill the Integration Parameter File in Quantitation Database Global page? 3. Also, after did what you said, the Concentration and Response could not be shown automatically.
@saudesaude3522
7 жыл бұрын
Hi Friend , Let me ask , what the purpose of the last way of calibtarion ? thanks , I ve got one of this Mass in my lab and I dont have clue how to do this ,
@agilent
7 жыл бұрын
Autoquant setup occasionally misses small peaks in the chromatogram. The last way permits these missed peaks to be added to the calibration table.
@saudesaude3522
7 жыл бұрын
espcially when two or more peack coeluitation , on mass itis possible to run in SIM mode in oder to exclude 3 drungs tht have same RT time ? thanks
@agilent
7 жыл бұрын
Yes
@prakashg7066
5 жыл бұрын
Super sir, Thank you so much ...
@saudesaude3522
7 жыл бұрын
for exemplo , if use head space and my sample has 1 grama where I put in this amount ?
@agilent
7 жыл бұрын
The sequence editor has a HELP button. Push it and scroll to the bottom of the Sequence Dialog Help topic, then select the “Sequence table columns” link to read about the “Amt.” and “Tot Amt.” columns. Then use the Tools\ Add/Remove columns to customize the user interface.
@thomasmammen4630
6 жыл бұрын
Which Version of MSD is it?
@agilent
6 жыл бұрын
The Enhanced Data Analysis software can be used with the 5973, 5975 and 5977 MSD.