Hey man! Great tip(s)! I've been listening your videos for a while now. Thank you for the content. I'm from Montreal, Canada. This is how I would make sure to solve the problem : 1 - Replace the seal - test 2 - Inspect isoplate interior for debris and to see the state of the inside probe bracket 3 - Inspect microswitch/ pimple/ probe / metal adapter 4 - Try another cassette - test 5 - Play with the angle of the cassette/ left-right while its pressurizing 6 - There is 2 big allen key screws on the side of the isoplate (left-right) that can be adjusted (not supposed too but we try that sometimes and it helps) If the probe bracket is in good shape (in/out) : 7 - Use the blue RPI (octagon) aligner without the cassette seal to make sure the probe and the pimple are well positioned to fit the cassette. (In your case the Statim seems fairly recent so I don't think its necessary). 8 - You can also start a cycle while the screws are not tightened to adjust the bracket accordingly to the cassette pressurizing sound to find the sweet spot (you can only tight 3 screws while doing so because of the live power under the boiler). If nothing works by trying to adjust the probe bracket and if its in bad shape : 9 - Install a new probe bracket (possibly a new probe as well) and adjust it to the cassette. If it's still not completing cycles and its leaking (cycle faults - water under the Statim / drops and vapor near the handle) : 10 - New cassette (Last resort - we had a customer who dropped his cassette on the floor. The bottom part was bent and we had to order a complete new one - problem solved). Lots of fun! Hope it helps you and the community!
@DentalEquipmentRepairChannel
Жыл бұрын
Wow! Incredible advice! Thank you so much. A true expert. 🙌
@maynorflores7167
3 ай бұрын
I just wanted to say that the video was absolutely amazing! Thanks for sharing it. By the way, I could use some advice. I have a STATIM G4 5000, and whenever it is operating, I hear a whistling sound. I'm not sure where the problem might be. Any tips on what could be causing this?
@DentalEquipmentRepairChannel
3 ай бұрын
Have you checked/replaced the gasket? It could be that or the alignment of the cassette "clamshell". Also look down the lower cassette half edges like you are sighting in a rifle and make sure it is absolutely straight and not bent in. If it is, slightly bend it out so it seals better against the seal.
@FL-gg4dq
3 ай бұрын
What ended up happening😊
@DentalEquipmentRepairChannel
3 ай бұрын
I re-adjusted the cassette probe bracket again. After running a little, I found that it only whistled when it was starting up. Once it reached temp, it held pressure. It was strange.