Hi Dr. Arvind Thank you so much for your helpful content I have a question if you would help me please, in the calculation of residual solvents (ethanol for example) do we include the density or relative density of ethanol in the equation? And if so, what will be the unit of density in this case? Thank you so much in advance.
@drarvindmittal7353
3 жыл бұрын
Hello Randa Gamal, Nice to hear you. As explained in the video, one should avoid putting in density while calculating weights of solvents, for the obvious reason of change in density and volume with temperature. I have explained the technique of weighing the solvents. One more point, which I would like to emphasize is that results are reported on ppm (parts per million) basis which are truly expressed if they are reported in ug/g basis (microgram per gram basis). From your query, I can conjecture that you intend to use micro-pipette for weighing of solvents, which is practically not correct. Hope this shall help.
@rahulchowdary9355
Жыл бұрын
how to calculate ppm for in solution foam?
@AjaySharma-jn4sr
3 жыл бұрын
Good Presentation
@Parthasarathivvc
2 жыл бұрын
In calculation part =sample area/std area×dilution factor× 2mL/samplewt×purity of solvent ×10000. Doubt was that 2mL was std solution taken in his vial or diluent +sample hss vial sir?
@drarvindmittal7353
2 жыл бұрын
Dear Sarathi, the volume of blank, standard and sample has to be the same. Only then we can expect equivalent concentration of analytes. If we are taking 2 mil as blank, then 2 mL of standard and 2 mL of diluent (as blank) shall be taken along with the sample in HSS vial. For more clarity, try to understand procedure. 1 vial blank (2mL diluent), 6 vials of standard containing 2 mL each of standard solution and 2 vials of sample+ 2 mL diluent in each sample vial. Hope this will clear the doubt.
@sivajinerella
3 жыл бұрын
Thank you sir
@achyutjagdhane9683
2 жыл бұрын
thanks a lot sir
@ankushthakur213
3 жыл бұрын
Hello Sir, I want to know how we select sample concentration or sample size in RS e.g. in some cases 200 mg/ml or in some cases 500mg/ 2ml
@drarvindmittal7353
3 жыл бұрын
Sample amount/size depends on the ICH limits of the solvents and cost of the sample. If-only Class-3 solvents are of interest, whose limits are generally 0.5% (5000 ppm), you may opt for lesser amounts. However, if Class-1 solvents are to be determined along with other solvents, then you will have to choose bigger amount, in order to increase impurity's response. Another factor is cost of the sample! In case of HPAPIs or oncological substances, price is one of the limiting factor.
@ankushthakur213
3 жыл бұрын
@@drarvindmittal7353 Thanks Sir
@pinaldesai4559
2 жыл бұрын
Sir, What is possible cause for vial skipping in HSS GC. Related to configure related
@karthick38208
Жыл бұрын
Sensor issue due to dust in G1888
@karthick38208
Жыл бұрын
Improper vial crimping in 7697 has and vial grip error
@parmjeetsingh8646
3 жыл бұрын
Sir ,i want to contact u to some help regarding residual solvent ...can be possible sir?
@drarvindmittal7353
3 жыл бұрын
You can reach me at arvind.mittal@albacasta.co.in
@anjaneyaprasad4241
3 жыл бұрын
please upload gc calibration procedure
@bhaskarlakavath
3 жыл бұрын
Sir how to derivation of methoxy amaine hydrochloride in how elluted peak can u inform sir
@drarvindmittal7353
3 жыл бұрын
Have you tried Schotten-Baumann reaction? Try to use benzoyl chloride in pyridine.